63) Transfer 1ml from LG-MHCB2 (right) to a new, clean small vial
„LG-CB". USE A CLEAN PIPETTE.
64) Place filled 1ml LG-CB vial into the adapter. Place the light shield
(PLSp-LS-1) on top and press TEST.
65) The determined result is immediately displayed. Press the "OK" button
to save the result together with date and time.
B) Measurement procedure for B) Cou TEST
1) Eluate filtered particles by shaking for 2 minutes.
2) Press OK to start 02:00 minutes countdown.
3) Press OK to start 15:00 minutes countdown.
4) Press OK to start 05:00 minutes countdown.
5) Press OK to start 03:00 minutes countdown.
6) Press OK to start 10:00 minutes countdown.
7) Press OK to start 03:00 minutes countdown.
8) Press OK to start 03:00 minutes countdown.
9) Press OK to start 03:00 minutes countdown.
10) Press OK to start 03:00 minutes countdown.
11) Press OK to start 02:00 minutes countdown.
12) Press OK to start 05:00 minutes countdown.
13) Place filled 1ml LG-CB vial into the adapter. Place the light shield
(PLSp-LS-1) on top and press ZERO.
14) Place filled 1ml LG-CB vial into the adapter. Place the light shield
(PLSp-LS-1) on top and press TEST.
Continued...
Name on device: 147-Legionella-liq
(147)
Legionella
Liquid
4
4
4
4
4
4
4
4
4
4
4
4
4
4
!
Possible messages in the line below the measurement value:
"low! / high! / good!":
Assessment of the measurement value relative to the ideal range defined by you.
"Overrange! / underrange!":
Measurement value for this parameter is outside the value range of this method.
Continued...
4
60 - 1000000 cfu/test
Содержание 1.0 MULTITEST
Страница 1: ......
Страница 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Страница 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Страница 33: ...left open for technical reasons ...
Страница 276: ...243 243 Tablet Note s Conversion to Na SO 2 3 Continued Name on device 53 Sulphite LR tab 53 Sulphite LR 0 5 mg l SO 3 ...
Страница 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Страница 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Страница 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Страница 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Страница 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Страница 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...