67
67
0 - 500 mg/l
(Pt-Co)
Name on device: 107-Colour
10 ml
24 mm
(107)
Colour
Select either:
A) „True colour“
B) „Apparent colour“
A) Measurement procedure for „True colour“
1.) Unscrew the two halves of the filter holder and place one filter circle
(0.45µ) onto the base section.
2.) Screw the two parts together again, ensuring the O ring is correctly
located.
3.) Fill a clean luer-lock syringe (20 ml) with deionised water.
4.) Connect the syringe to the filter holder and discharge the syringe to
waste.
5.) Disconnect the syringe from the filter holder and refill syringe with
destilled water.
6.) Fill a clean luer-lock syringe (20 ml) with deionised water.
7.) Connect the syringe to the filter holder and discharge the syringe to
waste.
8.) Disconnect the syringe from the filter holder and refill syringe with
destilled water.
9.) Fill a clean luer-lock syringe (20 ml) with deionised water.
10.) Connect the syringe to the filter holder and discharge the syringe to
waste.
11.) Disconnect the syringe from the filter holder and refill syringe with
destilled water.
(*)
Syringe
10
9
8
7
6
5
4
3
2
1
ml
Filter
0.45µm
(*)
Filter paper
(*)
Filter adapter
Continued...
The individual steps of the measurement procedure are shown on the display.
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
!
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5".
Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
Содержание 1.0 MULTITEST
Страница 1: ......
Страница 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Страница 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Страница 33: ...left open for technical reasons ...
Страница 276: ...243 243 Tablet Note s Conversion to Na SO 2 3 Continued Name on device 53 Sulphite LR tab 53 Sulphite LR 0 5 mg l SO 3 ...
Страница 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Страница 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Страница 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Страница 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Страница 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Страница 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...