Manganese VLR
Tablet
0 - 0.030 mg/l
(Mn)
Manganese VLR N°1
(TbsHMGNS1VLR...)
Manganese VLR N°2
(TbsPMGNS2VLR...)
10 ml
Name on device: 161-Manganes-VLR-tab
24 mm
1) Fill 10ml of the sample water into a clean 24mm cuvette.
2) Close the lid, place the cell into the photometer (use caution to the
arrow on the cell) and press ZERO.
3) Remove the cell and open the lid.
4) Add 1 Manganese VLR N°1 tablet to your sample.
5) Crush the tablet with a clean stirring rod until the tablet has completely
dissolved.
6) Add 1 Manganese VLR N°2 tablet to your sample.
7) Crush the tablet with a clean stirring rod until the tablet has completely
dissolved.
8) Close the lid. Shake the cell gently to mix the liquid.
9) Place the cell into the device. Again, please use caution to the arrow on
the cell. Than press TEST.
10) After the lapse of a 20:00 minutes countdown the determined result is
displayed.
Continued...
4
4
The individual steps of the measurement procedure are shown on the display.
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
!
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5".
Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
One-Time-Zero:
Measure new ZERO
Use last ZERO
For the sake of precision select "Use last ZERO" only
if the stored ZERO value is used with the
same cuvette and the same water sample for the test!
!
If after switching on the device you have already performed the
ZERO (baseline) adjustment for the same type, the following
message is displayed:
-> Measure new ZERO
= continue with step 2)
-> Use last ZERO
= continue with step 4)
1
2
3
4
Содержание 1.0 MULTITEST
Страница 1: ......
Страница 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Страница 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Страница 33: ...left open for technical reasons ...
Страница 276: ...243 243 Tablet Note s Conversion to Na SO 2 3 Continued Name on device 53 Sulphite LR tab 53 Sulphite LR 0 5 mg l SO 3 ...
Страница 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Страница 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Страница 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Страница 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Страница 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Страница 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...