72
72
Tablet
0.05 - 5 mg/l
(Cu)
Continued...
8) Remove the lid again and add a "Copper N° 2 Photometer" tablet into
the same sample you have just used for determining "fCu" (free copper).
9) Crush the tablet with a clean stirrer until it has completely dissolved.
10) Replace the lid on the cuvette, swirl it carefully and insert it into the
sampling chamber of the PrimeLab. Align the arrow on the front of the
cuvette.
Press TEST.
11) The total result is displayed immediately, divided in
"fCu" = "free copper“
"cCu" = "combined copper“
"tCu" = "total copper"
To scroll through the values please used the arrow up and down buttons.
The result is saved when button 4 is pressed.
Press button 4 (OK) to continue the measurement to determine the total copper
content (tCu) and later also the combined copper (cCu). If this is not desired the
measurement procedure can be terminated by pressing "HOME". In this case only
the valule for free copper (fCu) is stored.
!
2
3
4
4
(18)
Copper
(free/combined/total)
Name on device: 18-Copper-tab
!
Possible messages in the line below the measurement value:
"low! / high! / good!":
Assessment of the measurement value relative to the ideal range defined by you.
"Overrange! / underrange!":
Measurement value for this parameter is outside the value range of this method.
Continued...
Содержание 1.0 MULTITEST
Страница 1: ......
Страница 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Страница 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Страница 33: ...left open for technical reasons ...
Страница 276: ...243 243 Tablet Note s Conversion to Na SO 2 3 Continued Name on device 53 Sulphite LR tab 53 Sulphite LR 0 5 mg l SO 3 ...
Страница 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Страница 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Страница 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Страница 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Страница 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Страница 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...