131
131
Liquid
50 - 500 mg/l
2+
(Fe )
Name on device: 149-Iron-Oil-liq
TM-reagent-S
(TM149reagS)
TM-reagent-E
(TM149reagE)
TM-reagent-I
(TM149reagI)
TM-reagent-C
(TM149reagC)
(149)
Iron in Oil
Filter
0.45µm
(*)
(*)
(*)
Syringe
10
9
8
7
6
5
4
3
2
1
ml
Filter paper Filter adapter
The individual steps of the measurement procedure are shown on the display.
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
!
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5".
Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
Continued...
1) Open the lid of a clean (new) beaker.
2) Add 3ml of TM-reagent-E into the beaker.
3) Add 1ml of TM-reagent-C into the beaker.
4) Add 5ml of TM-reagent-I into the beaker.
5) Close the lid of the beaker and shake/swirl beaker until reagents are
mixed homogenously.
6) Place 3ml-vial-adapter into the PrimeLab.
7) Fill a clean luer-lock syringe with exactly 3ml of the water-part of your
sample.
8) Empty syringe into a clean/new 3ml vial.
9) Place 3ml vial into the adapter and apply light shield.
10) Press ZERO.
11) Remove light protection shield and 3ml vial.
12) Open the lid of a clean (new) beaker
4
One-Time-Zero:
Measure new ZERO
Use last ZERO
For the sake of precision select "Use last ZERO" only
if the stored ZERO value is used with the
same cuvette and the same water sample for the test!
!
If after switching on the device you have already performed the
ZERO (baseline) adjustment for the same type, the following
message is displayed:
-> Measure new ZERO
= continue with step 10)
-> Use last ZERO
= continue with step 11)
1
2
3
4
beaker
Adapter
Содержание 1.0 MULTITEST
Страница 1: ......
Страница 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Страница 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Страница 33: ...left open for technical reasons ...
Страница 276: ...243 243 Tablet Note s Conversion to Na SO 2 3 Continued Name on device 53 Sulphite LR tab 53 Sulphite LR 0 5 mg l SO 3 ...
Страница 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Страница 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Страница 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Страница 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Страница 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Страница 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...