132
132
Liquid
50 - 500 mg/l
2+
(Fe )
Name on device: 149-Iron-Oil-liq
(149)
Iron in Oil
13) Add precisely 0.3ml of Oil-sample to the beaker.
14) Add 3ml of TM-reagent-S into the beaker.
15) Add 3ml of TM-reagent-E into the beaker.
16) Add 1ml of TM-reagent-C into the beaker.
17) Add 5ml of TM-reagent-I into the beaker.
18) Tightly close the lid and shake beaker vigorously for 2 minutes.
19) Press TEST to start 02:00 minutes countdown.
20) Let beaker rest for 10:00 minutes. Press TEST to start 10:00 minutes
countdown.
21) Shake beaker vigorousely for 30 seconds.
22) Let beaker rest for 00:30 minutes. Press TEST to start 00:30 minutes
countdown.
23) Wait for the water to separate from Oil.
24) Fill a clean luer-lock syringe with exactly 3ml of the water-part of your
sample.
25) Connect syringe to a one-way-0.45µ filter and discharge 3ml into a
clean 3ml vial.
26) Place 3ml vial into the adapter and apply light shield.
27) Press TEST.
28) Press button 4 to save the result together with date and time in the
device.
Note(s)
-> Use cuation to use exact amounts of sample as well as of reagents
as indicated in the test procedure.
-> Always use new and clean vials and filter equipment. Do NOT re-use.
-> Keep 3ml-vials clean and free of fingerprints and dirt to avoid mis-
readings.
4
4
!
Possible messages in the line below the measurement value:
"low! / high! / good!":
Assessment of the measurement value relative to the ideal range defined by you.
"Overrange! / underrange!":
Measurement value for this parameter is outside the value range of this method.
4
4
4
Содержание 1.0 MULTITEST
Страница 1: ......
Страница 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Страница 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Страница 33: ...left open for technical reasons ...
Страница 276: ...243 243 Tablet Note s Conversion to Na SO 2 3 Continued Name on device 53 Sulphite LR tab 53 Sulphite LR 0 5 mg l SO 3 ...
Страница 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Страница 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Страница 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Страница 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Страница 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Страница 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...