Name on device: 160-Hydrocarbons
1) Fill a cleaned cell (24mm) with 2.5ml of your sample and 7.5ml of
deionised water
2) Close the lid. Shake the cell gently to mix the liquid.
3) Open the lid.
4.) Place the cell into the PrimeLab (use caution to the arrow on the cell).
Do NOT close the lid.
5.) Place TURBIDITY-ADAPTER on top of the open cuvette.
6.) Press TEST.
(160)
Hydrocarbons
4
The individual steps of the measurement procedure are shown on the display.
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
!
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5".
Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
!
Note(s)
->
Interferences:
- Hazy test sample
- The presence of excessive detergents
Interpretation:
0 = PASSED
OR = FAILED
10 ml
24 mm
PLSp-PIP10ml
PLSp-PIP10ml-tips
PLSp-ADP-TRB
Содержание 1.0 MULTITEST
Страница 1: ......
Страница 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Страница 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Страница 33: ...left open for technical reasons ...
Страница 276: ...243 243 Tablet Note s Conversion to Na SO 2 3 Continued Name on device 53 Sulphite LR tab 53 Sulphite LR 0 5 mg l SO 3 ...
Страница 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Страница 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Страница 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Страница 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Страница 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Страница 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...