266
266
(112)
Turbidity-NTU
0.02 - 1000
(NTU)
10 ml
Name on device: 112-Turbidity-NTU
24 mm
Adapter
PLSp-PIP10ml
PLSp-PIP10ml-tips
PLSp-ADP-TRB
Only for calibration!
Continued...
The individual steps of the measurement procedure are shown on the display.
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
!
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5".
Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
PLSp-RefTRB
10 ml
0.5
NTU
10 ml
10
NTU
10 ml
1000
NTU
1) Mix the sample water thoroughly, rinse cuvette -to be used for the
following measurement- several times and fill it with exactly 10ml of the
sample water. For this, you HAVE TO use the 10ml pipette which came
with your Turbidity Kit.
2) Insert the cuvette WITHOUT LID into the PrimeLab sampling chamber.
Align the arrow on the front of the cuvette.
3) Place Turbidity-Adapter on top of the open cuvette which already is
placed into the PrimeLab measurement chamber.
4) Press TEST.
5) The determined result is immediately displayed. Press button 3 to
convert the result to different units (*). Press button 4 to save the result
together with date and time.
3
4
4
!
Possible messages in the line below the measurement value:
"low! / high! / good!":
Assessment of the measurement value relative to the ideal range defined by you.
"Overrange! / underrange!":
Measurement value for this parameter is outside the value range of this method.
!!! If low values (<20 NTU) are expected, we recommend to let the water
sample (in the vial) rest for at least 05:00 minutes before presing TEST.
As an alternative, you can also continue to repeat measurement in steps of
01:00 minute. The lowest value displayed can be taken as a result.
Содержание 1.0 MULTITEST
Страница 1: ......
Страница 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Страница 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Страница 33: ...left open for technical reasons ...
Страница 276: ...243 243 Tablet Note s Conversion to Na SO 2 3 Continued Name on device 53 Sulphite LR tab 53 Sulphite LR 0 5 mg l SO 3 ...
Страница 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Страница 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Страница 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Страница 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Страница 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Страница 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...