Hardness-Calcium HR
Tablet
Calcium Hardness Photometer
(TbsPCH...)
10 ml
24 mm
1) Fill 10ml
distilled water
into a clean
24mm cuvette.
2) Add one "Calcium Hardness Photometer" tablet to the same cuvette.
3) Crush the tablet with a clean stirrer until it has completely dissolved.
4) Replace the screw lid on the cuvette and insert it into the PrimeLab
sampling chamber. Align the arrow on the front of the cuvette. Press
ZERO.
5) Wait for the lapse of a 2 minute countdown.
6) Remove the cuvette from the chamber and unscrew the lid.
7) Now add 2ml of your sample water, so that with 12ml the cuvette is
completely full.
8) Replace the lid on the cuvette and swirl this back and forth 5 times.
9) Replace the cuvette in the sampling chamber in the PrimeLab. Make
sure the arrow on the front of the cuvette is aligned. Press the TEST
button.
10) The determined result is immediately displayed. Press button 3 to
convert the result to different units (*). Press button 4 to save the result
together with date and time.
10
9
8
7
6
5
4
3
2
1
ml
This test method does not offer the OTZ (One Time Zero) function!
4
4
3
4
50 - 1000 mg/l
(CaCO )
3
Name on device: 09-Hard-Cal-HR-tab
Continued...
The individual steps of the measurement procedure are shown on the display.
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
!
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5".
Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
!
Possible messages in the line below the measurement value:
"low! / high! / good!":
Assessment of the measurement value relative to the ideal range defined by you.
"Overrange! / underrange!":
Measurement value for this parameter is outside the value range of this method.
Содержание 1.0 MULTITEST
Страница 1: ......
Страница 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Страница 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Страница 33: ...left open for technical reasons ...
Страница 276: ...243 243 Tablet Note s Conversion to Na SO 2 3 Continued Name on device 53 Sulphite LR tab 53 Sulphite LR 0 5 mg l SO 3 ...
Страница 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Страница 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Страница 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Страница 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Страница 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Страница 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...