124
124
0 - 10 mg/l
2+/3+
(Fe
)
(29)
Iron (MR)
Powder
Continued...
Note(s)
-> (*) Special accessories required / not included as standard equipment!
-> If the sample needs to be filtered (dissolved iron) it must be ensured that
after filtration there are no more undissolved iron parts in the sample. If you
are unsure, please repeat the filtration.
-> Very alkaline and acidic water samples must be adjusted to a pH value
between 3 and 5 before commencing the measurement.
-> The measurement is not influenced by undissolved powder.
-> If the water samples contains visible rust the reaction time must be 5
minutes. In this case wait for 2 minutes before initiating the step "TEST“
Name on device: 29-Iron-MR-pow
!
Possible messages in the line below the measurement value:
"low! / high! / good!":
Assessment of the measurement value relative to the ideal range defined by you.
"Overrange! / underrange!":
Measurement value for this parameter is outside the value range of this method.
Содержание 1.0 MULTITEST
Страница 1: ......
Страница 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Страница 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Страница 33: ...left open for technical reasons ...
Страница 276: ...243 243 Tablet Note s Conversion to Na SO 2 3 Continued Name on device 53 Sulphite LR tab 53 Sulphite LR 0 5 mg l SO 3 ...
Страница 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Страница 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Страница 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Страница 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Страница 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Страница 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...