Troubleshooting with adapters / Cleaning the device
A 3
Cleaning the device
Clean your PrimeLab preferably without using detergents.
Use special care in the area of the sampling chamber (where the cuvettes are
inserted). There you will find two round openings behind which there is a
transparent plastic pane, and behind that the LED resp. the sensor.
If these areas are not clean this will have a negative influence on the measurement
precision.
It is recommended to check these two openings regularly for soiling and clean them
with a cotton swab dipped in a little clear water.
Do not exert any pressure on the openings and the panes as the panes might
be pushed out, which would render the PrimeLab device unusable.
Also make sure that the two openings are completely dry after cleaning, without any
residues and are lint-free.
A 3
Troubleshooting with adapters
During calibration and / or during the measurement, the following error messages
might be displayed on the PrimeLab:
Error Message:
Adapter not recognized
Possible causes:
Battery depleted
Adapter incorrectly (eg diagonally) placed
Wrong adapter used (eg PTSA instead Turbidity)
Adapter-lense (bottom) or optical path in the sample
chamber
(PrimeLab) dirty or wet
Error message:
Measurement failed
Possible causes:
Battery depleted
Adapter incorrectly (eg diagonally) placed
The identified water sample does not match the turbidity
curve
Wrong adapter used (eg PTSA instead of Turbidity)
Adapter-lense (bottom) or optical path in the sample
chamber
(PrimeLab) dirty or wet
Содержание 1.0 MULTITEST
Страница 1: ......
Страница 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Страница 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Страница 33: ...left open for technical reasons ...
Страница 276: ...243 243 Tablet Note s Conversion to Na SO 2 3 Continued Name on device 53 Sulphite LR tab 53 Sulphite LR 0 5 mg l SO 3 ...
Страница 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Страница 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Страница 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Страница 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Страница 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Страница 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...