1) Fill 5ml distilled water into a clean 24mm cuvette.
2) Add precisely 5ml sample water to the same cuvette.
3) Replace the screw lid on the cuvette and insert it into the PrimeLab
sampling chamber. Align the arrow on the front of the cuvette. Press
ZERO.
4) Remove the cuvette from the chamber and unscrew the lid.
5) Add 2 drops "PL Urea 1" liquid reagent to the test liquid in the cuvette.
6) Replace the lid on the cuvette and swivel it carefully to mix the solution
well.
7) Add 1 drop "PL Urea 2" liquid reagent to the test liquid in the cuvette.
8) Replace the lid. Swivel the cuvette to mix the liquid and the reagent
until this is competely dissolved. Press the TEST button and wait for the
end of the 5 minute countdown. The cuvette does not have to be inside
the sampling chamber at this time.
9) Add a "Ammonia N° 1 Photometer" tablet to the water sample in the
cuvette.
Continued...
4
4
10 ml
24 mm
The individual steps of the measurement procedure are shown on the display.
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
!
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5".
Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
One-Time-Zero:
Measure new ZERO
Use last ZERO
For the sake of precision select "Use last ZERO" only
if the stored ZERO value is used with the
same cuvette and the same water sample for the test!
!
If after switching on the device you have already performed the
ZERO (baseline) adjustment for the same type, the following
message is displayed:
-> Measure new ZERO
= continue with step 3)
-> Use last ZERO
= continue with step 5)
1
2
3
4
Ammonia N° 1
(TbsHAM1...)
Ammonia N° 2
(TbsPAM2...)
PL Urea 1
(PL30Urea1)
PL Urea 2
(PL10Urea2)
0.2 - 5.0 mg/l
((NH ) CO)
2 2
(150)
Urea (HR)
Name on device: 150-UreaHR-tab-liq
Liquid
+Liquid
Tablet
Содержание 1.0 MULTITEST
Страница 1: ......
Страница 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Страница 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Страница 33: ...left open for technical reasons ...
Страница 276: ...243 243 Tablet Note s Conversion to Na SO 2 3 Continued Name on device 53 Sulphite LR tab 53 Sulphite LR 0 5 mg l SO 3 ...
Страница 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Страница 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Страница 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Страница 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Страница 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Страница 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...