TEST-9
Always conduct baseline (zero) measurements with the same cuvette
used for the subsequent test. Always make sure that the triangular
marking on the cuvette is aligned with the triangle on the front of the
sampling chamber on the device. There are always small differences
between cuvettes (tolerances due to production).
The device must be acclimatised to the ambient temperature. Great
differences between the device temperature and that of the environment
can lead to the formation of condensation obstructing the optical system,
which in turn will lead to incorrect measurements.
The sampling chamber must be free from water or humidity, otherwise
there will be the risk of damage to the electronics inside the device.
Please calibrate your PrimeLab on a regular basis (at least 2 weekly)
as described in SET-4 of these instructions to obtain the best
possible measurement results.
PrimeLab must remain on a flat surface while testing as otherwise
the LED light will not pass correctly through the sample water which
will lead to incorrect results.
Continued...
TEST-9
!
Important notes for accurate testing
Содержание 1.0 MULTITEST
Страница 1: ......
Страница 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Страница 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Страница 33: ...left open for technical reasons ...
Страница 276: ...243 243 Tablet Note s Conversion to Na SO 2 3 Continued Name on device 53 Sulphite LR tab 53 Sulphite LR 0 5 mg l SO 3 ...
Страница 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Страница 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Страница 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Страница 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Страница 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Страница 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...