Note(s)
-> Alkalinity of more than 250 mg/l CaCO affect the measurement and the
3
color development. Neutralize to pH 6-7 with 1 N Hydrochloric Acid before
performing the test.
-> Acidity of more than 150 mg/l CaCO affect the measurement and the color
3
development. Neutralize to pH 6-7 with 1 N Sodium Hydroxide before
performing the test.
-> Bromine at all levels, Ozone (if >1.5 mg/l) and chlorine (if >6 mg/l) as well as
iodine will interfere with the reagent reaction and will be part of the displayed
value. Peroxides may interfere as well.
-> Organic chloramines may interfere. Monochloramines lead to higher
readings (~0.1 mg/l at 3 mg/l monochloramines after 1 minute).
-> If oxidized manganese or oxidized chromium is present in the sample,
sample needs to be pre-treated.
-> Hardness levels above 1000 mg/l CaCO affect the measurement
3
-> Extreme pH-values of the sample need to be corrected to pH 6-7 before
measuring the sample.
0.00 - 5.00 mg/l
(ClO )
2
Name on device: 130-Chl-Diox-pp
(130)
Chlorine Dioxide
Liquid
+Powder
Содержание 1.0 MULTITEST
Страница 1: ......
Страница 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Страница 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Страница 33: ...left open for technical reasons ...
Страница 276: ...243 243 Tablet Note s Conversion to Na SO 2 3 Continued Name on device 53 Sulphite LR tab 53 Sulphite LR 0 5 mg l SO 3 ...
Страница 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Страница 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Страница 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Страница 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Страница 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Страница 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...