5
5
(121)
Alkalinity-M HR
Tablet
0 - 500 mg/l
(CaCO )
3
Name on device: 121-Alka-M-HR-tab
Continued...
Note(s)
-> (*) Conversion to HCO-3, °dH, °eH, °fH, mmol (KS4.3), mval
-> In order to obtain as precise a result as possible it is essential that the water
sample has a volume of exactly 10ml.
-> Extend the listed measurement range to 400 - 1000 mg/l by diluting your
water sample as follows: 1:1 = 5 ml of sample water plus 5 ml of distilled
water and continue with the test procedure. The test result displayed on the
screen needs to be multiplied by 2.