Phosphate -ortho- (HR)
0 - 100 mg/l
(PO )
4
Liquid
Continued...
10) After the lapse of a 10 minute countdown the determined result is
displayed. Press button 3 to convert this into various measurement
units (**).
11) Press button 4 to save the result together with date and time in the
device.
3
4
Name on device: 47-Phosphat-HR-liq
!
Possible messages in the line below the measurement value:
"low! / high! / good!":
Assessment of the measurement value relative to the ideal range defined by you.
"Overrange! / underrange!":
Measurement value for this parameter is outside the value range of this method.
Note(s)
-> (*) Special accessories required / not included as standard equipment!
-> (**) Conversion to P, P O
2
5
-> (***) Filter process is only needed in case of any suspended insoluble
phosphate expected in your water sample (applicable for boiler water
testing)
-> With this procedure ortho-phosphate ions are detected. Other phosphates
must therefore be converted to ortho-phosphates before the test is begun.
-> The pH value of the sample water should be between 6 and 7 pH.
-> The following contents of substances in the sample water can - at the
respective concentration - falsify the measurement results:
Chromium > 100 mg/l
Copper > 10 mg/l
Iron > 100 mg/l
Nickel > 300 mg/l
Zinc > 80 mg/l
Содержание 1.0 MULTITEST
Страница 1: ......
Страница 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Страница 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Страница 33: ...left open for technical reasons ...
Страница 276: ...243 243 Tablet Note s Conversion to Na SO 2 3 Continued Name on device 53 Sulphite LR tab 53 Sulphite LR 0 5 mg l SO 3 ...
Страница 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Страница 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Страница 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Страница 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Страница 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Страница 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...