190
190
Tablet
Name on device: 98-Phenol-tab
(98)
Phenol
0 - 5 mg/l
(C H OH)
6
5
Notes
-> If your sample does contain copper, zinc, iron or manganese ions (up to 350
mg/l) add one Phenol CR tablet after ZERO. Crush and mix to dissolve.
-> Level of > 20 mg/l hydrogen peroxide interferere with this test and can
influence the reading.
-> High (free) chlorine levels (> 10 mg/l) interfere with this test and can
influence the reading.
-> Alkalinity above 150 mg/l CaCO as well as sulphite above 10 mg/l or more
3
than 2 mg/l sulphide will interfere with this test and can influence the reading.
-> Some organic keto-enol compounds can lead to high readings.
!
Possible messages in the line below the measurement value:
"low! / high! / good!":
Assessment of the measurement value relative to the ideal range defined by you.
"Overrange! / underrange!":
Measurement value for this parameter is outside the value range of this method.
Continued...
Содержание 1.0 MULTITEST
Страница 1: ......
Страница 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Страница 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Страница 33: ...left open for technical reasons ...
Страница 276: ...243 243 Tablet Note s Conversion to Na SO 2 3 Continued Name on device 53 Sulphite LR tab 53 Sulphite LR 0 5 mg l SO 3 ...
Страница 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Страница 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Страница 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Страница 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Страница 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Страница 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...