Tablet
9) Wait for the lapse of a 2 minute countdown.
10) Remove the lid, empty the cuvette completely and clean it thoroughly.
11) Fill 10ml sample water into a second clean 24mm cuvette.
12) Add a "Glycine" tablet to the water sample in the cuvette.
13) Crush the tablet with a clean stirrer until this is completely dissolved.
14) Replace the lid on the cuvette and swirl it carefully to mix the liquids
well.
15) Add a "DPD N° 1 Photomoter" tablet and one "DPD N° 3 Photometer"
tablet into the first empty and clean cuvette.
16) Crush the tablets with a clean stirrer to a fine powdery mass.
17) Fill in the prepared water from the first cuvette.
18) Replace the lid on the cuvette, swirl it carefully and insert it into the
sampling chamber of the PrimeLab. Align the arrow on the front of the
cuvette. Press TEST.
19) After the lapse of a 2 minute countdown the total result is displayed,
divided in
"
O3" = "Ozone“
"
tCl" = "total chlorine“
Scroll through the results using the buttons "arrow up" and "arrow down".
Press button 4 to save the result.
Continued...
4
4
Name on device: 37-Ozone-tab
(37)
Ozone
!
Possible messages in the line below the measurement value:
"low! / high! / good!":
Assessment of the measurement value relative to the ideal range defined by you.
"Overrange! / underrange!":
Measurement value for this parameter is outside the value range of this method.
0 - 5.4 mg/l
(O )
3
Содержание 1.0 MULTITEST
Страница 1: ......
Страница 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Страница 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Страница 33: ...left open for technical reasons ...
Страница 276: ...243 243 Tablet Note s Conversion to Na SO 2 3 Continued Name on device 53 Sulphite LR tab 53 Sulphite LR 0 5 mg l SO 3 ...
Страница 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Страница 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Страница 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Страница 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Страница 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Страница 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...