(24)
Hydrogen Peroxide (LR)
Name on device: 24-Hydr-Per-LR-tab
Note(s)
-> If the measuring water contains further reducing chemicals (e.g. chlorine,
active oxygen, bromine etc.) this will also be detected and is part of the
result.
-> When preparing the measurement procedure it is important to avoid any
Hydrogen Peroxide escaping, which can happen during pipetting and
shaking the sample. The measurement should be performed directly after
sampling.
-> It is important that the measurement devices to be used have not been
cleaned with household detergent, as this could greatly reduce the
measurement. To prevent any contamination the cuvette, the cuvette lid and
the stirrer should be stored in a 0.1 g/l sodium hypochlorite solution for one
hour and then thoroughly rinsed with distilled water.
-> The reagent used in this procedure buffers the pH value of the sample water
in the range between 6.2 and 6.5 pH. If your sample water is very alkaline
or acidic this must be adjusted to a pH range between 6 and 7 by the
addition of 0.5 mol/l sulphuric acid or resp. 1 mol/l caustic soda before the
reagent is added.
-> Water values outside the above measurement range defined for this
parameter and this measurement procedure / this reagent can lead to
incorrect measurements. For the measurement of higher hyd. peroxide
values please select the respective matching procedure.
0 - 3.8
(H O )
2
2
mg/l
Содержание 1.0 MULTITEST
Страница 1: ......
Страница 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Страница 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Страница 33: ...left open for technical reasons ...
Страница 276: ...243 243 Tablet Note s Conversion to Na SO 2 3 Continued Name on device 53 Sulphite LR tab 53 Sulphite LR 0 5 mg l SO 3 ...
Страница 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Страница 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Страница 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Страница 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Страница 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Страница 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...