37
37
Continued...
Tablet
8) Remove the lid again and add a "DPD N° 3 Photometer" tablet into the
same sample you have just used for determining "free chlorine".
9) Crush the tablet with a clean stirrer until it has completely dissolved.
10) Replace the lid on the cuvette, swirl it carefully and insert it into the
sampling chamber of the PrimeLab. Align the arrow on the front of the
cuvette.
Press TEST.
11) After the lapse of a 2 minute countdown the total result is displayed,
divided in
"fCl" = "free chlorine“
"cCl" = "combined chlorine“
"tCl" = "total chlorine".
Scroll through the results using the buttons "arrow up" and "arrow down".
Press button 4 to save the result.
After pressing the button 4 (OK) the measurement is continued to determine the
total chlorine (tCl) and later also bonded chlorine (cCl). If this is not required the
measurement can be termianted by pressing "HOME". IN this case only the value
for free chlorine (fCl) is stored.
!
Continued...
2
3
4
4
0.00 - 8.00 mg/l
(Cl )
2
(11)
Chlorine
(free/combined/total)
Name on device: 11-Chlorine-tab
!
Possible messages in the line below the measurement value:
"low! / high! / good!":
Assessment of the measurement value relative to the ideal range defined by you.
"Overrange! / underrange!":
Measurement value for this parameter is outside the value range of this method.
Содержание 1.0 MULTITEST
Страница 1: ......
Страница 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Страница 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Страница 33: ...left open for technical reasons ...
Страница 276: ...243 243 Tablet Note s Conversion to Na SO 2 3 Continued Name on device 53 Sulphite LR tab 53 Sulphite LR 0 5 mg l SO 3 ...
Страница 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Страница 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Страница 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Страница 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Страница 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Страница 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...