188
188
Tablet
Name on device: 159-PTT-tab
(159)
Permanganate Time Test
Continued...
Note(s)
->
after adding the PTT tablet, sample is highly sensitive to light,
air and temperature. Do NOT open the vial after PTT tablet has been
added and lid got screwed on and keep it at constant temperature of
15°C
->
Interferences: turbid and/or colored water samples (before adding
PTT tablet)
0 - 100 %A
(Absorption)
1st test / after 10 minutes
2nd test / after 20 minutes
3rd test / after 30 minutes
4th test / after 40 minutes
5th test / after 50 minutes
test number
result
action
Abort test.
Tank is still highly contaminated.
No further testing
Replace vial in the fridge
and
repeat test after
10 minutes
(step 16 by pressing REPEAT
button = 2nd from left).
Do
add another tablet!).
NOT
If asked, select "use last ZERO"
<29% PTT
>29% PTT
<29% PTT
>29% PTT
<29% PTT
>29% PTT
<29% PTT
>29% PTT
<29% PTT
>29% PTT
Abort test.
Test has passed
.
Tank is
clean to Methanol stan-
dards.
Ready for next cargo
Abort test.
Tank is still highly contaminated.
No further testing
Replace vial in the fridge and
repeat test (step 16 by pressing
„REPEAT button = 2nd from left)
after 10 minutes (do
add
NOT
another tablet!).
If asked, select "use last ZERO"
Abort test.
Tank is still contaminated.
No further testing
Replace vial in the fridge and
repeat test (step 16 by pressing
„REPEAT button = 2nd from left)
after 10 minutes (do
add
NOT
another tablet!).
If asked, select "use last ZERO"
Abort test.
Tank is still contaminated.
No further testing
Replace vial in the fridge and
repeat test (step 16 by pressing
„REPEAT button = 2nd from left)
after 10 minutes (do
add
NOT
another tablet!).
If asked, select "use last ZERO"
Abort test.
Tank is still contaminated.
No further testing
Result:
18% (PTT)
1
2
3
4
REPEAT
Содержание 1.0 MULTITEST
Страница 1: ......
Страница 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Страница 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Страница 33: ...left open for technical reasons ...
Страница 276: ...243 243 Tablet Note s Conversion to Na SO 2 3 Continued Name on device 53 Sulphite LR tab 53 Sulphite LR 0 5 mg l SO 3 ...
Страница 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Страница 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Страница 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Страница 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Страница 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...