12.) Fill a clean luer-lock syringe (20 ml) with deionised water.
13.) Connect the syringe to the filter holder and discharge the syringe to
waste.
14.) Disconnect the syringe from the filter holder and refill syringe with
destilled water.
15.) Fill a clean luer-lock syringe (20 ml) with deionised water.
16.) Connect the syringe to the filter holder and discharge the syringe to
waste, down to the 10 ml mark.
17.) Filter the remaining 10 ml of destilled water into a clean 24mm vial.
18.) Close the lid, place the cell into the photometer (use caution to the
arrow on the cell) and press ZERO.
19.) Remove the cell and open the lid.
20.) Disconnect the syringe from the filter holder and refill syringe with
SAMPLE water.
21.) Connect the syringe to the filter holder and discharge the syringe to
waste.
22.) Disconnect the syringe from the filter holder and refill syringe with
SAMPLE water.
23.) Connect the syringe to the filter holder and discharge the syringe to
waste.
24.) Disconnect the syringe from the filter holder and refill syringe with
SAMPLE water.
25.) Connect the syringe to the filter holder and discharge the syringe to
waste, down to the 10 ml mark.
26.) Filter the remaining 10 ml of SAMPLE water into a clean 24mm vial.
27.) Close the lid, place the cell into the photometer (use caution to the
arrow on the cell) and press TEST.
4
Name on device: 107-Colour
(107)
Colour
Continued...
Continued...
4
0 - 500 mg/l
(Pt-Co)
Содержание 1.0 MULTITEST
Страница 1: ......
Страница 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Страница 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Страница 33: ...left open for technical reasons ...
Страница 276: ...243 243 Tablet Note s Conversion to Na SO 2 3 Continued Name on device 53 Sulphite LR tab 53 Sulphite LR 0 5 mg l SO 3 ...
Страница 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Страница 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Страница 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Страница 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Страница 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Страница 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...