58
58
Continued...
11) Fill sample water into the same cuvette up to the 10ml level.
12) Add 3 drops "PL DPD 3 C" liquid reagent to the test liquid in the
cuvette.
13) Replace the lid on the cuvette, swivel it carefully and insert it in the
sampling chamber of the PrimeLab. Align the arrow on the front of the
cuvette. Press TEST.
14) Wait for the lapse of a 2 minute countdown.
15) Remove the cuvette from the chamber and unscrew the lid.
16) Add 3 drops "PL DPD Acidifying" liquid reagent to the test liquid in the
cuvette.
17) Replace the lid on the cuvette, swivel it carefully and insert it in the
sampling chamber of the PrimeLab. Align the arrow on the front of the
cuvette. Press TEST.
18) Wait for the lapse of a 2 minute countdown.
19) Remove the cuvette from the chamber and unscrew the lid.
20) Add 3 drops "PL DPD Neutralising" liquid reagent to the test liquid in
the cuvette.
21) Replace the lid on the cuvette, swivel it carefully and insert it in the
sampling chamber of the PrimeLab. Align the arrow on the front of the
cuvette. Press TEST.
22) The determined result is immediately displayed. Press the "OK" button
to save the result together with date and time.
4
!
Possible messages in the line below the measurement value:
"low! / high! / good!":
Assessment of the measurement value relative to the ideal range defined by you.
"Overrange! / Underrange!":
Measurement value for this parameter is outside the value range of this method.
Note(s)
-> DPD 1 A and DPD 1 B reagent MUST be added to the vial BEFORE
water sample is added to avoid mis-readings!
-> The range of the side measurements performed during the Chlorite tests
may be up to 8 mg/l whilst it is rather unlikely that the range of Chlorite can
be tested up to 8 mg/l.
4
4
0 - 8 mg/l
(Cl )
2
Name on device: 106-Chlorite-liq
(106)
Chlorite
Liquid
Содержание 1.0 MULTITEST
Страница 1: ......
Страница 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Страница 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Страница 33: ...left open for technical reasons ...
Страница 276: ...243 243 Tablet Note s Conversion to Na SO 2 3 Continued Name on device 53 Sulphite LR tab 53 Sulphite LR 0 5 mg l SO 3 ...
Страница 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Страница 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Страница 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Страница 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Страница 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Страница 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...