Operator’s Manual – OPTI CCA-TS
3-25
3 CUSTOMIZATION
To set up Ethernet communication:
•
Connect instrument to an active network.
•
Go to
<System -> Diagnostics -> Tests
-> Ethernet>
.
•
Verify that the test status is
<Connected>
and you have a valid IP address (not all
zeroes) (Fig. 3-43).
•
Go back to menu
<System -> Setup ->
Hardware -> Interface>
(Fig. 3-44).
•
Choose communication format
<ASCII>
or
<ASTM>
.
•
Change communication link from
<Serial>
to
<Ethernet>
.
•
Choose the
<Ethernet>
tab (Fig. 3-45).
•
Set
<TCP Port>
to what your data manager
specifies is your instrument port. Each
analyzer type may have a different port
specified.
•
Enter your computer’s host IP address into
the
<Host IP>
field.
•
Press
.
•
<Format>
(E1394),
<IP Type>
(Dynamic), and
<UDP Port>
fields should
remain at default.
6. Press
to return to the
<Setup>
screen
or
to return to
<Ready>
.
Fig. 3-43 Ethernet Test
Fig. 3-44 Interface
Fig. 3-45 Ethernet
Содержание OPTI CCA-TS
Страница 1: ...OPTI CCA TS Analyzer Operator s Manual ...
Страница 2: ......
Страница 4: ......
Страница 10: ......
Страница 18: ......
Страница 28: ......
Страница 40: ......
Страница 69: ......
Страница 71: ......
Страница 89: ......
Страница 105: ......
Страница 118: ......
Страница 155: ......
Страница 171: ......
Страница 173: ......
Страница 182: ......
Страница 190: ......
Страница 198: ......
Страница 206: ......
Страница 210: ......
Страница 214: ......
Страница 218: ......
Страница 226: ......
Страница 239: ......
Страница 253: ......
Страница 254: ...Operator s Manual OPTI CCA TS B 1 APPENDIX B MENU STRUCTURE APPENDIX B MENU STRUCTURE ...
Страница 255: ......
Страница 257: ......