Operator’s Manual – OPTI CCA-TS
7-23
7 DIAGNOSTICS AND TROUBLESHOOTING
7.2.17 Checking the Display
The purpose of this test is to check the proper
operation of the display.
From the
<Ready>
display, press
<System
Manager -> <Diagnostics>
.
•
Select the
<Tests>
tab.
•
Press
<Display>
to start the test (Fig. 7-30).
•
The display will turn red, green and blue.
If this is not the case, your display is defective
and needs to be replaced.
•
Press
to return to the
<System>
screen.
Fig. 7-30 Display Test
7.2.16 Checking the Pump Motor
The purpose of this test is to check the proper
functioning of the peristaltic pump motor.
From the
<Ready>
display, press
<System
Manager -> Diagnostics>
.
•
Select the
<Tests>
tab.
•
Press
<Pump>
to start the test (Fig. 7-28).
•
The pump will automatically step through all the
speeds used during normal operation (50 to 800
pps (pulses per second)) (Fig. 7-29) and return to
the
<Tests>
screen.
Fig. 7-28 Pump Motor Test
Fig. 7-29 Pump Speed
Speed: 800
Pump Test . . .
Speed: 800
Pump Test . . .
Содержание OPTI CCA-TS
Страница 1: ...OPTI CCA TS Analyzer Operator s Manual ...
Страница 2: ......
Страница 4: ......
Страница 10: ......
Страница 18: ......
Страница 28: ......
Страница 40: ......
Страница 69: ......
Страница 71: ......
Страница 89: ......
Страница 105: ......
Страница 118: ......
Страница 155: ......
Страница 171: ......
Страница 173: ......
Страница 182: ......
Страница 190: ......
Страница 198: ......
Страница 206: ......
Страница 210: ......
Страница 214: ......
Страница 218: ......
Страница 226: ......
Страница 239: ......
Страница 253: ......
Страница 254: ...Operator s Manual OPTI CCA TS B 1 APPENDIX B MENU STRUCTURE APPENDIX B MENU STRUCTURE ...
Страница 255: ......
Страница 257: ......