Cl-1
ANALYTES
CHLORIDE
Chloride (Cl
-
)
Clinical Significance
1
Chloride is an anion that exists predomininantly in extracellular spaces. It maintains cellular integrity
through its influence on osmotic pressure. It is also significant in monitoring acid-base balance and water
balance. In metabolic acidosis, there is a reciprocal rise in chloride concentration when the bicarbonate
concentration drops.
Decreased levels are found in severe vomiting, severe diarrhea, ulcerative colitis, pyloric obstruction,
severe burns, heat exhaustion, diabetic acidosis, Addison’s disease, fever and acute infections such as
pneumonia.
Increased levels are found in dehydration, Cushing’s syndrome, hyperventilation, eclampsia, anemia and
cardiac decompensation.
Measurement Principle
The Cl
-
ion optodes
are closely related to the more familiar Ion Selective Electrodes (ISEs).
The optodes use ion selective recognition elements (ionophores) similar to those used in ISEs, however
the ionophores are linked to fluorescent dyes instead of electrodes. These types of dyes have been
used since the 1970’s to visualize and quantify cellular ion levels in fluorescence microscopy and cell
counters
2
.
As the ion concentration increases, these ionophores bind larger amounts of ions and cause the
fluorescence intensity to increase or decrease, depending on the particular ion. Like the pH optode, the
ion optodes do not need a reference electrode, however, several of them do exhibit a small pH sensitivity
which is automatically compensated in the OPTI CCA-TS using the measured pH.
Measurement Range
Range
Resolution (Low/High)
Units
50 to 160 mmol/L
1/0.1
mmol/L
Standard Reference Cassette (SRC) Limit Values
LOW
NORMAL
HIGH
Units
80.0 ± 2
105.0 ± 2
130.0 ± 2
mmol/L
Содержание OPTI CCA-TS
Страница 1: ...OPTI CCA TS Analyzer Operator s Manual ...
Страница 2: ......
Страница 4: ......
Страница 10: ......
Страница 18: ......
Страница 28: ......
Страница 40: ......
Страница 69: ......
Страница 71: ......
Страница 89: ......
Страница 105: ......
Страница 118: ......
Страница 155: ......
Страница 171: ......
Страница 173: ......
Страница 182: ......
Страница 190: ......
Страница 198: ......
Страница 206: ......
Страница 210: ......
Страница 214: ......
Страница 218: ......
Страница 226: ......
Страница 239: ......
Страница 253: ......
Страница 254: ...Operator s Manual OPTI CCA TS B 1 APPENDIX B MENU STRUCTURE APPENDIX B MENU STRUCTURE ...
Страница 255: ......
Страница 257: ......