Operator’s Manual – OPTI CCA-TS
PO2-B-1
ANALYTES
P
O
2
(Dry Sensor)
PO
2
(Dry Sensor - B-Lac Cassette)
Clinical Significance
1
The
P
O
2
value of arterial blood is used to assess how well the body is able to absorb oxygen in the lungs.
Values below the normal arterial
P
O
2
(
arterial hypoxemia
) are usually caused by pulmonary, circulatory,
or respiratory abnormalities (e.g. bronchial obstruction, vascular problems, decrease in cardiac output,
increased oxygen demand, anatomical heart defect, low inspired O
2
content). Generally, O
2
levels
above 100 mmHg do not contribute significantly to the oxygen content since, with normal hemoglobin
concentrations, 80 - 100 mmHg,
P
O
2
provides a 97% saturation level, and a level greater than 100%
cannot be achieved.
Measurement Principle
The
P
O
2
optode measurement principle
is based upon luminescence quenching, first documented in the
1930’s
2
,
and commercially utilized to measure blood
P
O
2
in 1983
3
.
The relationship of luminescence to
P
O
2
is quantified by the Stern-Volmer equation,
I
0
/ I = 1 + kP
which describes how the fluorescence emission intensity “I” is reduced as the
P
O
2
“P”, is increased.
Unlike conventional electrochemical “Clark”
P
O
2
electrodes, the oxygen optode does not consume
oxygen molecules during the measurement.
The
P
O
2
partial pressure is influenced by the local barometric pressure, as dictated by Dalton’s law.
The OPTI CCA-TS incorporates a pressure transducer, which accurately tracks the local barometric
pressure and automatically compensates for it. The OPTI CCA-TS has been factory-calibrated to the
absolute barometric pressure.
Measurement Range
Range
Resolution (Low/High)
Units
10 to 700
1/0.1
mmHg
1.30 - 93.31
0.1/0.01
kPa
Standard Reference Cassette (SRC) Limit Values
LOW
NORMAL
HIGH
Units
60.0 ± 3
100.0 ± 3
170.0 ± 3
mmHg
8.00 ± 0.40
13.33 ± 0.40
22.66 ± 0.40 kPa
Содержание OPTI CCA-TS
Страница 1: ...OPTI CCA TS Analyzer Operator s Manual ...
Страница 2: ......
Страница 4: ......
Страница 10: ......
Страница 18: ......
Страница 28: ......
Страница 40: ......
Страница 69: ......
Страница 71: ......
Страница 89: ......
Страница 105: ......
Страница 118: ......
Страница 155: ......
Страница 171: ......
Страница 173: ......
Страница 182: ......
Страница 190: ......
Страница 198: ......
Страница 206: ......
Страница 210: ......
Страница 214: ......
Страница 218: ......
Страница 226: ......
Страница 239: ......
Страница 253: ......
Страница 254: ...Operator s Manual OPTI CCA TS B 1 APPENDIX B MENU STRUCTURE APPENDIX B MENU STRUCTURE ...
Страница 255: ......
Страница 257: ......