Ca-2
Operator’s Manual – OPTI CCA-TS
CALCIUM
ANALYTES
Interferences
The OPTI CCA-TS Ca
++
sensor does exhibit an interference from bisulfate and phenylacetic acid.
Heparin salts are the only acceptable anticoagulants. Other anticoagulants such as citrate, EDTA, oxalate,
and fluoride cause significant interferences to the electrolyte sensors.
The following exogenous interferents were quantified in tonometered plasma, showing interferences to
dyes which typically have short half-lives within the body before being metabolized by the liver.
Substance
amount Ca
++
change mmol/L
Sodium fluorescein
26 mg/dL
unstable
Cardio (indocyanine) green
0.5 mg/dL
+0.01
Methylene blue
25 mg/dL
unstable
Only clear, uncolored quality control materials, such as OPTI CHECK or OPTI CHECK PLUS brand
aqueous controls should be used with the OPTI CCA-TS system. Colored materials, including proficiency
testing materials, may interfere with the ion measurement, or fail to be properly aspirated.
Reproducibility
Typical Within-Run (
Swr
), Between-Day (
Sdd
) and Total (
ST
) Precision is determined from 1 run per
day with 2 replicates per run for 20 days on each of two OPTI CCA instruments. Ionized calcium values
are expressed in mmol/L.
Material
mean S
wr (CV%)
S
dd (CV%)
S
T (CV%)
OPTI CHECK PLUS
(aqueous control solution)
Level 1
1.26
0.001 (0.1)
0.007 (0.6)
0.007 (0.6)
Level 2
1.16
0.002 (0.2)
0.010 (0.9)
0.011 (0.9)
Level 3
0.79
0.001 (0.1)
0.005 (0.7)
0.005 (0.7)
Serum
0.64
0.001 (0.1)
0.01 (1.8)
0.01 (1.8)
Reduced Bovine Hemoglobin
0.81
0.001 (0.2)
0.007 (0.9)
0.007 (0.9)
Solution
All specific performance characteristics tests were run with default instrument calibration and after
normal recommended equipment quality control checks were performed (see Operator’s Manual).
Specimens at each level were analyzed in replicates of two for 20 days. The within-run and
between-day standard deviations were calculated by the analysis of variance method.
Содержание OPTI CCA-TS
Страница 1: ...OPTI CCA TS Analyzer Operator s Manual ...
Страница 2: ......
Страница 4: ......
Страница 10: ......
Страница 18: ......
Страница 28: ......
Страница 40: ......
Страница 69: ......
Страница 71: ......
Страница 89: ......
Страница 105: ......
Страница 118: ......
Страница 155: ......
Страница 171: ......
Страница 173: ......
Страница 182: ......
Страница 190: ......
Страница 198: ......
Страница 206: ......
Страница 210: ......
Страница 214: ......
Страница 218: ......
Страница 226: ......
Страница 239: ......
Страница 253: ......
Страница 254: ...Operator s Manual OPTI CCA TS B 1 APPENDIX B MENU STRUCTURE APPENDIX B MENU STRUCTURE ...
Страница 255: ......
Страница 257: ......