Operator’s Manual – OPTI CCA-TS
PCO2-3
ANALYTES
P
CO
2
Linearity
3
Wherever possible, linearity for the OPTI CCA measurement has been established against reference
materials or methods.
P
CO
2
linearity is established against values determined on whole blood
tonometered to gravimetrically prepared gases with CO
2
concentrations certified to 0.03% absolute by the
manufacturer, and measured on three OPTI CCA systems.
Correlation
Slope
Intercept
Coefficient
Sy.x
Range
n
0.9681
2.148
0.99967
1.53
11 - 201
81
Correlation to Other Methods
3
OPTI CCA vs other pH/Blood Gas Instruments on whole blood in a typical setting
Excess blood aliquots from specimens collected for blood gas analyses were analyzed by both traditional
and non-traditional operators of blood gas equipment in hospital laboratories. The blood was analyzed on
the OPTI CCA after obtaining the requisite results from existing instrumentation used for these analyses
and operated and controlled following their established procedures.
Correlation
Comparative Method*
Slope
Intercept
Coefficient Sy.x
Range
n
Analyzer A (whole blood)
0.9751
1.623
0.9871
1.16
28 – 72
103
Analyzer B (whole blood)
0.9740
2.66
0.9937
1.12
24 – 92
173
Analyzer C (whole blood) 0.988 ± 0.022 0.807± 1.015
0.9750
2.584
23 – 81
105
Analyzer D (whole blood) 1.073 ± 0.011 -2.785 ± 0.521
0.9910
2.050
17 – 122
174
Analyzer E (whole blood) 1.067 ± 0.009 -4.41 ± 0.468
0.9936
1.817
22 – 120
183
* For more information on specific analyzers used, please contact OPTI Medical Technical Support.
References
1. Tietz, Norbert W., Ed.,
Clinical Guide to Laboratory Tests
, 2nd Ed., (Philadelphia: W.B.Saunders,
Co., 1990) p. 436
.
2. Vurek GG, Feustel PJ, Severinghaus JW,
A Fiber Optic
P
CO
2
Sensor
, Ann.Biomed.Eng. 11, p.499,
1983.
3. OPTI Medical.
Model equation for regression statistics is: [results of OPTI Analyzer] = slope(m)
[comparative method results] + intercept(b)
.
Содержание OPTI CCA-TS
Страница 1: ...OPTI CCA TS Analyzer Operator s Manual ...
Страница 2: ......
Страница 4: ......
Страница 10: ......
Страница 18: ......
Страница 28: ......
Страница 40: ......
Страница 69: ......
Страница 71: ......
Страница 89: ......
Страница 105: ......
Страница 118: ......
Страница 155: ......
Страница 171: ......
Страница 173: ......
Страница 182: ......
Страница 190: ......
Страница 198: ......
Страница 206: ......
Страница 210: ......
Страница 214: ......
Страница 218: ......
Страница 226: ......
Страница 239: ......
Страница 253: ......
Страница 254: ...Operator s Manual OPTI CCA TS B 1 APPENDIX B MENU STRUCTURE APPENDIX B MENU STRUCTURE ...
Страница 255: ......
Страница 257: ......