7-20
Operator’s Manual – OPTI CCA-TS
7 DIAGNOSTICS AND TROUBLESHOOTING
Fig. 7-18 Optics Test
Fig. 7-19 Remove SRC
7.2.13 Checking the Optics
Fig. 7-16 Optics Test
Fig. 7-17 Insert SRC
This option checks the output of the six optics
channels. This test is designed for trained service
personnel.
From the
<Ready>
display, press
<System
Manager>
and
<Diagnostics>
.
•
Select the
<Tests>
tab.
•
Press
<Optics>
to start the test (Fig. 7-16).
•
Insert an SRC and press
(Fig. 7-17).
•
The system will now check the optics
(Fig. 7-18).
•
At the completion of the test, a printout of the
results will be printed and the
<Remove SRC>
message will be displayed (Fig. 7-19).
Test in Process . . .
Optics Test . . .
Test in Process . . .
Optics Test . . .
Remove SRC
Optics Test . . .
Remove SRC
Optics Test . . .
Insert SRC
Optics Test . . .
OK
Cancel
Insert SRC
Optics Test . . .
OK
Cancel
Содержание OPTI CCA-TS
Страница 1: ...OPTI CCA TS Analyzer Operator s Manual ...
Страница 2: ......
Страница 4: ......
Страница 10: ......
Страница 18: ......
Страница 28: ......
Страница 40: ......
Страница 69: ......
Страница 71: ......
Страница 89: ......
Страница 105: ......
Страница 118: ......
Страница 155: ......
Страница 171: ......
Страница 173: ......
Страница 182: ......
Страница 190: ......
Страница 198: ......
Страница 206: ......
Страница 210: ......
Страница 214: ......
Страница 218: ......
Страница 226: ......
Страница 239: ......
Страница 253: ......
Страница 254: ...Operator s Manual OPTI CCA TS B 1 APPENDIX B MENU STRUCTURE APPENDIX B MENU STRUCTURE ...
Страница 255: ......
Страница 257: ......