7-12
Operator’s Manual – OPTI CCA-TS
7 DIAGNOSTICS AND TROUBLESHOOTING
The gas pressure is low.
•
Replace the gas bottle (See Section 6.5.1).
The temperature is out of range.
•
Wait for the analyzer to reach the correct
temperature.
•
Cycle the power if the analyzer does not go to
<Ready>
within a few minutes.
•
If the analyzer does not become
<Ready>
within a reasonable time, check the temperature
under
<System - Diagnostics>
.
Check that ambient temperature is within
operating specifications on page 2-1.
The battery voltage is low.
•
Operate the analyzer on AC power and/or
recharge the battery.