3-12
Operator’s Manual – OPTI CCA-TS
3 CUSTOMIZATION
3.3.2.5 Setting
Normal Ranges or Alarm Limits
This menu enables you to change both the limits
“name” as it appears on the printout and the limit
values themselves, for pH,
P
CO
2
and all other
measured parameters. These limit names can be
based on your hospital policy and may be selected
from the following - “Reference”, “Normal”,
“Physiologic”,” Alarm” or “Critical”.
A result that is outside the limits you define here
will be flagged with an up-arrow if high, or down-
arrow if low. A message is included on the printout
explaining each arrow, using the name selected here.
NOTE: When the patient temperature has been
changed, both the uncorrected and corrected
parameters will be checked against the
limit values programmed here and flagged
accordingly.
1. From the
<Ready>
display, select
<System Manager>
and
<Setup>
.
2. In the
<System Setup>
menu, press
<Normal
Ranges/Alarm Limits>
(Fig. 3-23).
3. On the
<Limits 1>
tab, select the limits range
name you wish to use (Fig. 3-24):
Reference
Normal
Physiologic
Alarm
Critical
4. Press
<Limits 2>
to advance to the next screen
(Fig. 3-25).
5. Select the parameter you want to change and
press
to enter the new limit value.
6. Press
to accept the new limit value.
Fig. 3-23 Select Normal Ranges
Fig. 3-24 Select Limits Name
Fig. 3-25 Enter Limits 2
Содержание OPTI CCA-TS
Страница 1: ...OPTI CCA TS Analyzer Operator s Manual ...
Страница 2: ......
Страница 4: ......
Страница 10: ......
Страница 18: ......
Страница 28: ......
Страница 40: ......
Страница 69: ......
Страница 71: ......
Страница 89: ......
Страница 105: ......
Страница 118: ......
Страница 155: ......
Страница 171: ......
Страница 173: ......
Страница 182: ......
Страница 190: ......
Страница 198: ......
Страница 206: ......
Страница 210: ......
Страница 214: ......
Страница 218: ......
Страница 226: ......
Страница 239: ......
Страница 253: ......
Страница 254: ...Operator s Manual OPTI CCA TS B 1 APPENDIX B MENU STRUCTURE APPENDIX B MENU STRUCTURE ...
Страница 255: ......
Страница 257: ......