Operator’s Manual – OPTI CCA-TS
7-21
7 DIAGNOSTICS AND TROUBLESHOOTING
7.2.14 Checking the RS232 Interface
The purpose of this test is to check for proper
functioning of the serial interface.
From the
<Ready>
display, press
<System
Manager -> Diagnostics>
.
•
Select the
<Tests>
tab.
•
Press
<RS232>
to start the test (Fig. 7-20).
•
It is important to have pins 2 and 3 (send and
receive - see Appendix A, p. A-6) shorted
together (Fig. 7-21).
•
Press
and the system will send out a
test string and check if it can be received.
•
The instrument will display a
<Pass>
or
<Fail>
message (Fig. 7-22).
•
Press
to return to the
<Tests>
screen.
Fig. 7-20 RS232 Interface
Fig. 7-21 Jumper Pins 2 to 3
Fig. 7-22 Interface Test Pass
Jumper pins 2 to 3.
RS232 Test . . .
OK
Jumper pins 2 to 3.
RS232 Test . . .
OK
RS232: PASS
RS232 Test . . .
OK
RS232: PASS
RS232 Test . . .
OK
Содержание OPTI CCA-TS
Страница 1: ...OPTI CCA TS Analyzer Operator s Manual ...
Страница 2: ......
Страница 4: ......
Страница 10: ......
Страница 18: ......
Страница 28: ......
Страница 40: ......
Страница 69: ......
Страница 71: ......
Страница 89: ......
Страница 105: ......
Страница 118: ......
Страница 155: ......
Страница 171: ......
Страница 173: ......
Страница 182: ......
Страница 190: ......
Страница 198: ......
Страница 206: ......
Страница 210: ......
Страница 214: ......
Страница 218: ......
Страница 226: ......
Страница 239: ......
Страница 253: ......
Страница 254: ...Operator s Manual OPTI CCA TS B 1 APPENDIX B MENU STRUCTURE APPENDIX B MENU STRUCTURE ...
Страница 255: ......
Страница 257: ......