7-16
Operator’s Manual – OPTI CCA-TS
7 DIAGNOSTICS AND TROUBLESHOOTING
7.2.7 Checking the Cooling Fan
The purpose of this test is to check for proper
functioning of the cooling fan.
From the
<Ready>
display, press
<System
Manager>
and
<Diagnostics>
.
•
Select the
<Control>
tab.
•
Press the
<On/Off>
button under
<Fan>
to start the test (Fig. 7-8).
•
When
<On>
is selected, you should feel the
draft of the fan by placing your hand over the
fan at the back side of the analyzer.
•
Press
to return to the
<System>
screen.
Fig. 7-8 Cooling Fan
7.2.6 Checking the Battery Voltage
This selection lets you check the battery voltage.
From the
<Ready>
display, press
<System
Manager>
and
<Diagnostics>
.
•
The second display in the
<Miscellaneous>
section shows the battery voltage (Fig. 7-7).
•
If the voltage is below 11.8V, the battery needs
to be recharged or may need replacement.
Fig. 7-7 Battery Voltage
Содержание OPTI CCA-TS
Страница 1: ...OPTI CCA TS Analyzer Operator s Manual ...
Страница 2: ......
Страница 4: ......
Страница 10: ......
Страница 18: ......
Страница 28: ......
Страница 40: ......
Страница 69: ......
Страница 71: ......
Страница 89: ......
Страница 105: ......
Страница 118: ......
Страница 155: ......
Страница 171: ......
Страница 173: ......
Страница 182: ......
Страница 190: ......
Страница 198: ......
Страница 206: ......
Страница 210: ......
Страница 214: ......
Страница 218: ......
Страница 226: ......
Страница 239: ......
Страница 253: ......
Страница 254: ...Operator s Manual OPTI CCA TS B 1 APPENDIX B MENU STRUCTURE APPENDIX B MENU STRUCTURE ...
Страница 255: ......
Страница 257: ......