7-24
Operator’s Manual – OPTI CCA-TS
7 DIAGNOSTICS AND TROUBLESHOOTING
7.2.18 Checking the Touch Screen
The purpose of this test is to check the proper
operation of the touch screen.
From the
<Ready>
display, press
<System
Manager -> Diagnostics>
.
•
Select the
<Tests>
tab.
•
Press
<Touch>
to start the test (Fig. 7-31).
•
Touch the screen and a dot should appear under
the touched location (Fig. 7-32).
•
If not, press
<Calibrate>
to perform a touch
calibration.
•
Press
to return to the
<System>
screen.
•
Using a finger, stylus or pointed object (e.g.
syringe adapter), touch the center of the
calibration mark as it moves around the screen
(Fig. 7-33).
NOTE: Do not use sharp objects, since they may
damage the screen.
•
When finished press
.
Fig. 7-31 Select Touch Test
Fig. 7-32 Perform Touch Test
Fig. 7-33 Perform Touch Calibration
Содержание OPTI CCA-TS
Страница 1: ...OPTI CCA TS Analyzer Operator s Manual ...
Страница 2: ......
Страница 4: ......
Страница 10: ......
Страница 18: ......
Страница 28: ......
Страница 40: ......
Страница 69: ......
Страница 71: ......
Страница 89: ......
Страница 105: ......
Страница 118: ......
Страница 155: ......
Страница 171: ......
Страница 173: ......
Страница 182: ......
Страница 190: ......
Страница 198: ......
Страница 206: ......
Страница 210: ......
Страница 214: ......
Страница 218: ......
Страница 226: ......
Страница 239: ......
Страница 253: ......
Страница 254: ...Operator s Manual OPTI CCA TS B 1 APPENDIX B MENU STRUCTURE APPENDIX B MENU STRUCTURE ...
Страница 255: ......
Страница 257: ......