Operator’s Manual – OPTI CCA-TS
Ca-1
ANALYTES
CALCIUM
Ionized Calcium (Ca
++
)
Clinical Significance
1
Calcium in blood is distributed as free calcium ions (50%); bound to protein, mostly albumin (40%); and
10% bound to anions such as bicarbonate, citrate, phosphate and lactate. However, only ionized calcium
can be used by the body in such vital processes as muscular contraction, cardiac function, transmission of
nerve impulses and blood clotting. The OPTI CCA-TS measures the ionized portion of the total calcium.
In certain disorders such as pancreatitis and hyperparathyroidism, ionized calcium is a better indicator for
diagnosis than total calcium.
Elevated calcium,
hypercalcemia
, may be present in various types of malignancy, and calcium
measurements may serve as biochemical markers. In general, while ionized calcium may be slightly more
sensitive, either ionized or total calcium measurements have about equal utility in the detection of occult
malignancy. Hypercalcemia occurs commonly in critically ill patients with abnormalities in acid-base
regulation and losses of protein and albumin, which gives a clear advantage to monitoring calcium status
by ionized calcium measurements.
Patients with renal disease caused by glomerular failure often have altered concentrations of calcium,
phosphate, albumin, magnesium and pH. Since these conditions tend to change ionized calcium
independently of total calcium, ionized calcium is the preferred method of accurately monitoring calcium
status in renal disease
2
.
Ionized calcium is important for diagnosis or monitoring of: hypertension management, parathyroidism,
renal diseases, malnutrition, kidney stones, multiple myeloma and diabetes mellitus.
Measurement Principle
The Ca
++
ion optodes
are closely related to the more familiar Ion Selective Electrodes (ISEs).
The optodes use ion selective recognition elements (ionophores) similar to those used in ISEs, however
the ionophores are linked to fluorescent dyes instead of electrodes. These types of dyes have been
used since the 1970’s to visualize and quantify cellular ion levels in fluorescence microscopy and cell
counters
3
.
As the ion concentration increases, these ionophores bind larger amounts of ions and cause the
fluorescence intensity to increase or decrease, depending on the particular ion. Like the pH optode, the
ion optodes do not need a reference electrode, however, several of them do exhibit a small pH sensitivity
which is automatically compensated in the OPTI CCA-TS using the measured pH.
Measurement Range
Range
Resolution (Low/High)
Units
0.2 to 3.0
0.01
mmol/L
Standard Reference Cassette (SRC) Limit Values
LOW
NORMAL
HIGH
Units
1.8 ± 0.1
1.1 ± 0.1
0.7 ± 0.1
mmol/L
Содержание OPTI CCA-TS
Страница 1: ...OPTI CCA TS Analyzer Operator s Manual ...
Страница 2: ......
Страница 4: ......
Страница 10: ......
Страница 18: ......
Страница 28: ......
Страница 40: ......
Страница 69: ......
Страница 71: ......
Страница 89: ......
Страница 105: ......
Страница 118: ......
Страница 155: ......
Страница 171: ......
Страница 173: ......
Страница 182: ......
Страница 190: ......
Страница 198: ......
Страница 206: ......
Страница 210: ......
Страница 214: ......
Страница 218: ......
Страница 226: ......
Страница 239: ......
Страница 253: ......
Страница 254: ...Operator s Manual OPTI CCA TS B 1 APPENDIX B MENU STRUCTURE APPENDIX B MENU STRUCTURE ...
Страница 255: ......
Страница 257: ......