8-6
Operator’s Manual – OPTI CCA-TS
8 OPERATING PRINCIPLES
8.5.2 Test Conditions
Sample Size:
a minimum of 125 µL (60µL for B60 cassette)
Sample Type:
heparinized whole
blood, serum and plasma
Sample Application:
syringe, capillary or ComfortSampler
Ambient Temperature:
10 - 32 ºC (50 – 90 ºF)
Relative Humidity:
5% to 95% (non-condensing)
Type of Measurement:
optical fluorescence (pH,
P
O
2
,
P
CO
2
, Na
+
, K
+
, Ca
++
) Cl
-
,
Glu, BUN (urea), Lac) and reflectance (tHb,
S
O
2
)
8.5.3 Input Values
Parameter
Range
Default Display Units
Resolution
Patient ID
15 alphanumeric characters
Blank
Operator ID
11 alphanumeric characters
Blank
Accession Number
12 alphanumeric characters
Blank
Patient temperature, T
14 to 44
37.0
0.1
°C
58 to 111
0.1
°F
Patient Sex
Male, female or ?
?
Hemoglobin type
adult or fetal
adult
DOB
MMM-DD-YYYY
Puncture Site
LR/RR/LB/RB/LF/RF/
LR
Cord/Scalp, where:
LR = Left Radial
RR = Right Radial
LB = Left Brachial
RB = Right Brachial
LF = Left Femoral
RF = Right Femoral
Cord = Cord
Scalp = Scalp
Bypass
Off Pump / On Pump
Off Pump
Содержание OPTI CCA-TS
Страница 1: ...OPTI CCA TS Analyzer Operator s Manual ...
Страница 2: ......
Страница 4: ......
Страница 10: ......
Страница 18: ......
Страница 28: ......
Страница 40: ......
Страница 69: ......
Страница 71: ......
Страница 89: ......
Страница 105: ......
Страница 118: ......
Страница 155: ......
Страница 171: ......
Страница 173: ......
Страница 182: ......
Страница 190: ......
Страница 198: ......
Страница 206: ......
Страница 210: ......
Страница 214: ......
Страница 218: ......
Страница 226: ......
Страница 239: ......
Страница 253: ......
Страница 254: ...Operator s Manual OPTI CCA TS B 1 APPENDIX B MENU STRUCTURE APPENDIX B MENU STRUCTURE ...
Страница 255: ......
Страница 257: ......