Operator’s Manual – OPTI CCA-TS
7-11
7 DIAGNOSTICS AND TROUBLESHOOTING
The gas cylinder is empty.
•
Replace the gas cylinder and press
.
If SRC QC lockout has been activated in
<Setup>
,
this message will be displayed if SRCs have not
been run within the specified time.
•
Press
and run SRCs.
The instrument has detected an internal error.
•
Discard the cassette.
•
Turn the power off, wait 30 seconds and then
turn the power back on.
Patient, QC and other databases were deleted.
•
Press
and the instrument will
reinitialize.
This error message will appear if you try to run
lactate cassettes if the lactate parameter has not been
set up on your analyzer.
•
Refer to section 7.2.20 for instructions to set up
the lactate parameter.
If QC lockout has been activated in
<Setup>
this
message will be displayed if controls have not been
run within the specified time.
•
Press
and run control materials.
The temperature is out of range during any kind of
measurement.
•
Press
and continue.
•
If the error message appears again, check the
temperature under
<System - Diagnostics>
.
STOP! - Low Gas.
Replace Now.
OK
STOP! - Low Gas.
Replace Now.
OK
STOP! - Need New SRC Data.
Must Run SRCs.
OK
STOP! - Need New SRC Data.
Must Run SRCs.
OK
STOP! - System Error.
Cycle Power for Reset.
STOP! - System Error.
Cycle Power for Reset.
STOP! - Memory Error.
Database Deleted.
OK
STOP! - Memory Error.
Database Deleted.
OK
STOP! - Lactate Setup Required.
Lactate Setup . . .
OK
STOP! - Lactate Setup Required.
Lactate Setup . . .
OK
STOP! - Need New QC Data.
Must Run Controls.
OK
STOP! - Need New QC Data.
Must Run Controls.
OK
STOP! - Temperature Out of Range.
OK
STOP! - Temperature Out of Range.
OK
Содержание OPTI CCA-TS
Страница 1: ...OPTI CCA TS Analyzer Operator s Manual ...
Страница 2: ......
Страница 4: ......
Страница 10: ......
Страница 18: ......
Страница 28: ......
Страница 40: ......
Страница 69: ......
Страница 71: ......
Страница 89: ......
Страница 105: ......
Страница 118: ......
Страница 155: ......
Страница 171: ......
Страница 173: ......
Страница 182: ......
Страница 190: ......
Страница 198: ......
Страница 206: ......
Страница 210: ......
Страница 214: ......
Страница 218: ......
Страница 226: ......
Страница 239: ......
Страница 253: ......
Страница 254: ...Operator s Manual OPTI CCA TS B 1 APPENDIX B MENU STRUCTURE APPENDIX B MENU STRUCTURE ...
Страница 255: ......
Страница 257: ......