7-22
Operator’s Manual – OPTI CCA-TS
7 DIAGNOSTICS AND TROUBLESHOOTING
7.2.15 Checking the Pump Flow
This option is designed to test the pump cartridge.
From the
<Ready>
display, press
<System
Manager>
and
<Diagnostics>
.
•
Select the
<Tests>
tab.
•
Press
<Flow>
to start the test (Fig. 7-23).
•
Insert a new cassette (Fig 7-24).
•
Close the SMC cover.
•
Wait for test results (Fig. 7-25).
•
The two numbers indicate the actual flow rates
clockwise and counter clockwise (Fig. 7-26).
If one of the two or both rates are out of range,
the test fails.
•
Repeat test or replace the pump cartridge, if
the test fails. See replacement instructions in
Chapter 6.
NOTE: It is possible that the test fails the first
time, even if the pump cartridge is working
correctly.
•
Remove the cassette (Fig. 7-27).
Fig. 7-23 Flow Test
Fig. 7-25 Flow Test in Process
Fig. 7-24 Insert Cassette
Fig. 7-26 Flow Test Fail
Fig. 7-27 Remove Cassette
Insert Cassette
FlowTest . . .
Cancel
Insert Cassette
FlowTest . . .
Cancel
Test in Process . . .
FlowTest . . .
Test in Process . . .
FlowTest . . .
FlowTest . . .
Flow: 4.48 99.9
FAIL: Replace pump
OK
FlowTest . . .
Flow: 4.48 99.9
FAIL: Replace pump
OK
Please Remove the Cassette
FlowTest . . .
Please Remove the Cassette
FlowTest . . .
Содержание OPTI CCA-TS
Страница 1: ...OPTI CCA TS Analyzer Operator s Manual ...
Страница 2: ......
Страница 4: ......
Страница 10: ......
Страница 18: ......
Страница 28: ......
Страница 40: ......
Страница 69: ......
Страница 71: ......
Страница 89: ......
Страница 105: ......
Страница 118: ......
Страница 155: ......
Страница 171: ......
Страница 173: ......
Страница 182: ......
Страница 190: ......
Страница 198: ......
Страница 206: ......
Страница 210: ......
Страница 214: ......
Страница 218: ......
Страница 226: ......
Страница 239: ......
Страница 253: ......
Страница 254: ...Operator s Manual OPTI CCA TS B 1 APPENDIX B MENU STRUCTURE APPENDIX B MENU STRUCTURE ...
Страница 255: ......
Страница 257: ......