3-10
Operator’s Manual – OPTI CCA-TS
3 CUSTOMIZATION
3.3.2.3 Selecting Which Parameters Are Blanked/Disabled
In the
<Measured Parameters>
menu you can
enable parameter blanking and disable certain
parameters from being reported on the analyzer.
1. From the
<Ready>
display, select
<System Manager>
and
<Setup>
.
2. In the
<System Setup>
menu, press
<Measured Parameters>
(Fig. 3-19).
3. Press
<Allow Blanking>
to allow parameter
blanking (Fig. 3-20).
If blanking is enabled, the user is prompted
to choose which measured parameter will be
disabled or removed from the record after each
patient sample measurement. If for example,
Ca
++
is disabled, this result will not appear in the
stored patient results or on the printout.
4. Select the parameters to be disabled.
5. Press
to save the settings.
6. Press
to return to the
<Setup>
screen
or
to return to
<Ready>
.
Fig. 3-19 Measured Parameters
Fig. 3-20 Select Allow Blanking
Содержание OPTI CCA-TS
Страница 1: ...OPTI CCA TS Analyzer Operator s Manual ...
Страница 2: ......
Страница 4: ......
Страница 10: ......
Страница 18: ......
Страница 28: ......
Страница 40: ......
Страница 69: ......
Страница 71: ......
Страница 89: ......
Страница 105: ......
Страница 118: ......
Страница 155: ......
Страница 171: ......
Страница 173: ......
Страница 182: ......
Страница 190: ......
Страница 198: ......
Страница 206: ......
Страница 210: ......
Страница 214: ......
Страница 218: ......
Страница 226: ......
Страница 239: ......
Страница 253: ......
Страница 254: ...Operator s Manual OPTI CCA TS B 1 APPENDIX B MENU STRUCTURE APPENDIX B MENU STRUCTURE ...
Страница 255: ......
Страница 257: ......