Operator’s Manual – OPTI CCA-TS
PO2-3
ANALYTES
P
O
2
Linearity
4
Wherever possible, linearity for the OPTI CCA measurement has been established against reference
materials or methods.
P
O
2
linearity is established against values determined on whole blood tonometered
to gravimetrically prepared gases with O
2
concentrations certified to 0.03% absolute by the manufacturer,
and measured on three OPTI CCA systems.
Correlation
Slope
Intercept
Coefficient
Sy.x
Range
n
0.9844
1.864
0.99974
4.52
21 – 500
68
Correlation to Other Methods
4
OPTI CCA vs other pH/Blood Gas Instruments on whole blood in a typical setting
Excess blood aliquots from specimens collected for blood gas analyses were analyzed by both traditional
and non-traditional operators of blood gas equipment in hospital laboratories. The blood was analyzed on
the OPTI CCA after obtaining the requisite results from existing instrumentation used for these analyses
and operated and controlled following their established procedures.
Correlation
Comparative Method*
Slope
Intercept
Coefficient Sy.x
Range
n
Analyzer A (whole blood)
0.9419
3.28
0.9875
18.27
36 - 563
103
Analyzer B (whole blood)
1.0192
-4.13
0.9969
4.10
34 - 291
173
Analyzer C (whole blood) 0.918 ± 0.087 8.083 ± 1.402
0.9954
8.032
29 – 407
105
Analyzer D (whole blood) 1.041 ± 0.006 -6.244 ± 0.931
0.9969
6.379
37 – 598
174
Analyzer E (whole blood) 0.993 ± 0.009 1.646 ± 0.893
0.9925
4.458
34 – 321
183
* For more information on specific analyzers used, please contact OPTI Medical Technical Support.
References
1. Tietz, Norbert W., Ed.,
Clinical Guide to Laboratory Tests
, 2nd Ed., (Philadelphia: W.B.Saunders,
Co., 1990) p. 436
.
2. Kautsky H,
Quenching of Luminescence by Oxygen
, Transactions Faraday Society 35, p.216, 1939
3. CDI,
3M Healthcare System 200 Extracorporeal Blood Gas Monitor
. See, for example, Lubbers DW,
Gehrich J, Opitz N,
Fiber Optics Coupled Flourescence Sensors for Continuous Monitoring of Blood
Gases in the Extracorporeal Circuit
, Life Supports Systems 4, p.94, 1986.
4. OPTI Medical.
Model equation for regression statistics is: [results of OPTI Analyzer] = slope(m)
[comparative method results] + intercept(b)
.
Содержание OPTI CCA-TS
Страница 1: ...OPTI CCA TS Analyzer Operator s Manual ...
Страница 2: ......
Страница 4: ......
Страница 10: ......
Страница 18: ......
Страница 28: ......
Страница 40: ......
Страница 69: ......
Страница 71: ......
Страница 89: ......
Страница 105: ......
Страница 118: ......
Страница 155: ......
Страница 171: ......
Страница 173: ......
Страница 182: ......
Страница 190: ......
Страница 198: ......
Страница 206: ......
Страница 210: ......
Страница 214: ......
Страница 218: ......
Страница 226: ......
Страница 239: ......
Страница 253: ......
Страница 254: ...Operator s Manual OPTI CCA TS B 1 APPENDIX B MENU STRUCTURE APPENDIX B MENU STRUCTURE ...
Страница 255: ......
Страница 257: ......