Operator’s Manual – OPTI CCA-TS
8-13
8 OPERATING PRINCIPLES
8.5.10 Interferences
Selected substances endogenous and exogenous to human blood were tested for interference in accord
with CLSI EP7-A2
5
.
These substances were selected on the basis of their optical absorbance or
fluorescence properties likely to affect the optical signal measured by the OPTI CCA-TS, or the optical
properties of the sensor measured by the analyzer. To cause interference to the optical sensors, the
substances must be highly mobile (low molecular weight) and highly colored, in order to penetrate
the optode membrane barriers quickly (within the 90 sec. measurement interval), and then strongly
absorb light or emit light of the proper color. To cause interference to the tHb and
S
O
2
reflectance
measurements, the substances must strongly absorb or scatter red or infrared light, relative to normal
whole blood.
The following substances were tested in whole blood at the CLSI-recommended test level or higher,
and showed no interference to any measured analyte, including blood gas, electrolytes, and tHb/
S
O
2
:
Bile Acids (30 µmol/dL)
Bilirubin (40 mg/dL)
Beta-Carotene (3.0 mg/dL)
Hemolysis (10%) During hemolysis K
+
is released from the blood cells thereby
increasing the measured K
+
. In the same manner, protein released
from the cells binds ionized Ca
++
and decreases the concentration.
While an accurate value is reported, it will reflect the actual changes
caused by hemolysis.
Lipemia (equivalent to 3000 mg/dL triglycerides)
Elevated white blood cell count (30,000 WBC/
µ
L)
The following substances were tested in plasma at the CLSI-recommended test level or higher, and
showed no interference to blood gas and electrolyte analytes:
Coumadin (Warfarin) (12 mg/dL)
Dicumarol (Dicoumarin) (11 mg/dL)
Procain (Novacaine) (13 mg/dL)
Acetaminophen (Paracetamol) (20 mg/dL)
The OPTI CCA-TS system was evaluated for the interference of sample temperature on measurement
(iced samples). No measurable sensitivity to sample temperature was found.
5
Clinical and Laboratory Standards Institute (CLSI). Interference Testing in Clinical Chemistry;
Approved Guideline - 2nd Edition. CLSI document EP7-A2. CLSI, Wayne, PA, 2005
Содержание OPTI CCA-TS
Страница 1: ...OPTI CCA TS Analyzer Operator s Manual ...
Страница 2: ......
Страница 4: ......
Страница 10: ......
Страница 18: ......
Страница 28: ......
Страница 40: ......
Страница 69: ......
Страница 71: ......
Страница 89: ......
Страница 105: ......
Страница 118: ......
Страница 155: ......
Страница 171: ......
Страница 173: ......
Страница 182: ......
Страница 190: ......
Страница 198: ......
Страница 206: ......
Страница 210: ......
Страница 214: ......
Страница 218: ......
Страница 226: ......
Страница 239: ......
Страница 253: ......
Страница 254: ...Operator s Manual OPTI CCA TS B 1 APPENDIX B MENU STRUCTURE APPENDIX B MENU STRUCTURE ...
Страница 255: ......
Страница 257: ......