108
Memory Bank Designation Write: WM
Section 3-19
When Setting Unit is 0.1
1.
The response block for Alarm Temperature Read (R%) does not include
read data if the end code of the response block is other than 00.
2.
Refer to
2-4 End Codes
.
Communications Example
In this example, the E5ZE is operated with Alarm Temperature Read (R%)
under the following conditions.
Unit no.:
1
Memory Bank no.:
2
Control Point:
3
Setting Unit:
1
Alarm 1 Temperature: 50
°
C
Command
Response
3-19 Memory Bank Designation Write: WM
Function
This command is used to designate the memory banks at a control point that
are used for temperature control.
1.
Memory Bank Designation Write (WM) cannot be used if the memory bank
designation method is set to contact input designation.
2.
Memory Bank Designation Write (WM) cannot be used at a control point
being auto
-
tuned.
3.
When the E5ZE is turned ON, the memory bank numbers previously se-
lected through communications will be effective.
Setting Data Range
@ R %
s
Terminator
*
Alarm set value
FCS
x16
x16
Unit
Read data
Header code
End code
x16
x16
x10
x10
x10
x10
Or minus sign (-)
x10
Decimal point
1
1
1
0
0
0
3
2
-1
@ 0 1 R % 2 3 0 0 3 7
s
Alarm 1
*
@ 0 1 R % 0 0 0 0 5 0 3 3
s
Normal end
50
°
C
*
Default
0000 for all control points
Setting data
0000 to 0007