85
3
3.
MAINTENANCE
3.1
Troubleshooting
In the event that the ICE-1000 does not work correctly, then troubleshoot with the following table
before contacting NIDEK or your authorized distributor.
If the symptom cannot be corrected with the above actions, contact NIDEK or your authorized distributor.
Symptom
Action
The ICE-1000 does not work at all.
• Confirm that the power cord is properly connected.
• Confirm that the voltage applied to the outlet is within the range specified.
• Check the fuses. If the fuses are blown, replace them with new ones.
See ““3.2 Replacing Fuses” (Page 86)” for fuse replacement.
The measured data of the lens is
unstable.
• Wipe dirt off on the lens table (grid plate).
• Keep the instrument away from direct illumination.
The following error message is
displayed during startup of the
blocker.
• Optical Initialization Error
• A grid detection error occurred during power-up.
• Remove foreign substances and soiling from the lens table and turn the blocker
on again.
• Detach the lens table and turn the blocker on again.
When the blocker is restarted successfully, attach the lens table.
One of the following message is
displayed:
• X Movement Error
• Y Movement Error
• R Movement Error
• Reverse Movement Error
• Block Movement Error
• Clamp Error
• Upper Pin Operation Error
• Lower Pin Operation Error
• The blocking mechanism is abnormal. After checking the error message, shut
down the ICE-1000 and contact NIDEK or your authorized distributor.
Содержание ICE-1000
Страница 1: ...INTELLIGENT BLOCKER Model ICE 1000 OPERATOR S MANUAL...
Страница 8: ...VI...
Страница 114: ...104 SPECIFICATIONS AND ACCESSORIES Standard Configuration 5 3 2 Optional accessories Barcode scanner RS 232C cable...