42
OPERATING PROCEDURES
:
Setting Processing Conditions and Entering Layout Data
The range of the FPD value is 30.00 to 99.50 mm in increments of 0.01 mm.
Pressing the CE button clears the entered value. When no value is entered or the value is
not changed, this button closes the numeric keypad.
When the traced data includes the frame data (both-rim
tracing), the FPD value is automatically entered
together with the traced data.
The DBL (width between the nasal ends of the left and
right frames) can also be entered instead of the FPD.
Pressing the FPD button displays the pup-up menu. Press the DBL button to enter the DBL
value.
The range of the DBL value is 1.00 to 30.00 mm in increments of 0.01 mm.
2) PD (or 1/2PD)
Press the PD (or 1/2PD) numeric button to display the numeric keypad. Enter any value and press the
Ent button.
The range of the PD value is 30.00 to 99.50 mm in increments of 0.01 mm.
The monocular PD (1/2PD) can also be entered
instead of the PD.
It is possible to enter the distance from the
bridge center to each eye separately.
PD
FPD
Height of the
optical center
¨ǽ©
ᵁᴿᵉ
1/2PD
(R)
1/2PD
(L)
FPD
Height of
optical center (
ǽ
)
PD