98
CONNECTION AND SETTINGS
:
Connection
4.2.2
Setting the ICE-1000
Set the ICE-1000 according to the system configuration as:
*1: Set the parameter on the Network screen of the ICE-1000.
*2: Never change the factory setting of Mac Address that had been set uniquely for each individual
ICE-1000 before shipment.
*3: Set these parameters in accordance with the network to which the ICE-1000 is connected.
*4: When the NIDEK-produced PC server software is used, set the host port No. to “55555”. When the
other server software is used, set the host port No. to the port No. of the used software.
Parameter
Mini LAB
RS232
Entry LAB
Extended LAB
Edger port
LAN port
ID No.
ID No. of the ICE-1000: 32
–
ID No. of the ICE-1000:
(Any number except for the Server
No.)
Sever No.
ID No. of the ICE-1000: 32
–
ID No. of the server:
32
Baud rate
38400
–
38400
–
Mac Address*
1
–
Never change the
factory setting.*
2
–
Never change the factory setting.*
2
IP Address*
1
–
Set the parameter
uniquely for each
individual ICE-1000.*
3
–
Set the parameter uniquely for
each individual ICE-1000.*
3
Host IP Address*
1
–
–
–
Set the parameter uniquely for
each individual ICE-1000.*
3
(IP Address of the server)
Sub Net Mask*
1
–
Set the parameter
uniquely for each
individual ICE-1000.*
6
–
Set the parameter uniquely for
each individual ICE-1000.*
3
Gate Way*
1
–
–
–
Set the parameter uniquely for
each individual ICE-1000.*
3
Host Port No*
1
–
–
–
55555*
4