88
MAINTENANCE
:
Adjusting the Cup Holder
3.4
Adjusting the Cup Holder
If the plunger inside the cup holder becomes loose or worn out, the cup holder may not hold the lens
cup securely.
To tighten the plunger, insert the hexagonal-head wrench into the hole at the front of the cup holder,
and turn the plunger so that the cup holder can securely hold the lens cup, and that the lens cup can
be detached or inserted smoothly.
Turn the plunger clockwise to increase the
holding power of the lens cup.
3.5
Cleaning
When the covers or panels of the ICE-1000 becomes dirty, wipe with a soft cloth. For severe dirt,
immerse the cloth in a neutral detergent, wring well, and wipe. Finally wipe with a dry and soft cloth.
3.6
List of Replacement Parts
* After replacing consumables, restock them.
Plunger
CAUTION
• Never use an organic solvent such as paint thinner.
This may ruin the surface of the instrument.
• Lightly wipe the exterior of the LCD touch screen with a soft cloth. Do not press the
LCD touch screen using an object with a hard tip. In addition, keep magnetic objects
away from the LCD touch screen.
Scratches or failure of the LCD touch screen may result.
• Never use a sponge or cloth soaked in water.
The water may leak into the inside of the instrument and cause instrument failure.
Part name
Part number
Note
Fuse
80402-02150
T 1.6 A 250 V 5 X 20 mm.