79
OPERATING PROCEDURES
:
Blocking
2
{
Error messages when of the blocking error
The cause of the blocking error is displayed with the error message.
Error message
Action
Blocking position is out of range.
The blocking position is out of the blockable range.
In the modes other than Manual, move the lens position or switch the
layout mode between active and passive.
In Manual mode, switch the layout mode between active and passive.
Optical center is out of range.
The optical center detected in Single: Optical center mode is out of
the blockable range.
Move the lens so that it is within the blockable range.
There is undetermined layout data.
Determine the layout data.
Blocking protection is ON.
The lens of either eye side is going to be blocked with the same
(already-used) traced data.
To block the lens, press the LEFT or RIGHT button for two seconds or
more. The blocking protection is released.
Detection Error
The lens is not detected or the measurement is unstable. Check the
following:
• The lens is placed on the lens table.
• The lens and lens table are clean.
• No interference lights
Axis Error (5 degrees)
The segment axis of the bifocal lens exceeds 5°.
Rotate the lens so that the segment axis is within 5°.
Please clamp a lens.
The lens is not fully tightened with the lens clamp.
Press the Lens clamp button to tighten the lens.