IV
:
After Use
Maintenance
CAUTION
• If the instrument will not be used, turn the power off and close the tracing unit
cover.
Settled dust may affect the measurement accuracy.
• Do not yank the power cord to disconnect it from a wall outlet but hold the plug.
This can damage the metal core of the cord and may result in electric shock, short cir-
cuit, or fire.
• Occasionally clean the prongs of the power plug with a dry cloth.
If dust settles between the prongs, the dust will collect moisture, and short circuit or
fire may occur.
• If the instrument will not be used for a long time, disconnect the power cord from
the wall outlet.
Fire may occur.
• Maintain the surrounding temperature and humidity at the following range during
transport and storage of the packed instrument.
Environmental conditions:
Temperature: –25 to +70 ºC
Humidity: 10 to 95% (Non-condensing)
A place with low dust
A place not exposed to direct sunlight
• To transport the instrument, use the special packing materials to protect the
instrument from shock or impact.
Excessive vibration or impact to the instrument may cause malfunction.
CAUTION
• Be sure to perform check before and after use. It is recommended to have regular
checks every two years.
Regular checks must be performed by qualified personnel. Ask NIDEK or your autho-
rized distributor for details.
• When performing maintenance work, secure a sufficient maintenance space. (See
Page V.)
Maintenance work in an insufficient space may result in injury.
• Use the specified fuses to replace the old ones.
If not, fire may result.