33
OPERATING PROCEDURES
:
Tracing
2
{
Goggle type frame tracing
When the frames are sharply warped, the stylus may come off the groove. In such a case, fasten only
the one frame between the rim clips and perform single-eye tracing.
This tracing is available in both left and right frames. The procedure for tracing the right eye frame is
described here.
1
Set the right frame in the tracing unit so that
the frame becomes level.
Do not fasten the left frame between the rim
clips.
Hold the left temple to maintain the right
frame level.
2
Start tracing.
Press the
Right-eye tracing button
.
3
Remove the frames from the tracing unit.
4
Call up the traced data on the display.
Press the Tracer button on the LCD touch screen.
The lens shape is displayed.
• The rim clips fasten the right frame. However, do not release the left frame until tracing is
completed.
When the frame moves, tracing is not performed accurately.
• The FPD (Frame Pupil Distance) can not be measured in goggle type frame tracing.
Enter the FPD value manually on the layout screen.